Wanbo Group Instruments · Authorized Test & Measurement Supplier
Communication failures usually come from wrong menu settings, wrong cables or unconfigured PC software. This guide covers RS232, USB and LAN debugging in three steps.
Introduction: As automated testing grows, PCs increasingly control instruments. ITECH offers rich interfaces and remote-control functions, but communication can fail. This guide helps you troubleshoot quickly.
ITECH products support RS232, RS485, USB, LAN, GPIB and CAN. Confirm which interface your model has, select it in the menu and configure parameters. For LAN, put the instrument and PC on the same subnet (different last IP octet, same subnet mask and gateway) so you can ping it.
Use the right cable for each interface (USB cable, network cable, etc.). Note that models such as IT8500 and IT6800 with optional communication interface only have a TTL port on the rear and require an optional ITECH adapter (IT-E121/E122/E123) to connect to a PC.
ITECH provides free PC software; download IT9000 and the driver from the official site's share center. You can also control via SCPI (most models) or FRAME format (e.g. IT8500; IT8500+ supports both).
Serial: use a serial tool, pick the correct COM port and baud rate, end each command with a newline and send ASCII; send the remote command first. For hex, multiply voltage by 1000 and current by 10000 before converting.
USB: use NI MAX; if no port appears, update the COM-port driver in Device Manager.
LAN: set the mode to LAN, configure IP/gateway/subnet, ping the instrument, then use NI MAX.