Use the SIGLENT SVA1015X tracking generator to test an unknown attenuator, including normalization calibration and measurement steps.
Attenuators are elements of an electrical circuit that decrease the power delivered to the next stage. An ideal attenuator has a flat frequency response, decreasing the amplitude of all input signals equally regardless of frequency.
In this note, we use the tracking generator (TG) function of a SIGLENT SVA1015X Spectrum Analyzer to test an unknown attenuator. The TG provides a sine wave signal with a known amplitude across a range of frequencies, and we measure the response of the device-under-test (DUT).
Normalization minimizes measurement errors added by cables, adapters and connectors. Connect all cabling and adapters in place of the DUT using a through-adapter, run the sweep, then activate normalization so the result subtracts the cabling effects point-by-point. The normalized curve should show as a flat line at 0 dB before adding the DUT.
To avoid damaging connectors, grip the adapter and spin the hex cable end to tighten — spinning the adapter can quickly damage connectors.