Wanbo Group Instruments · Authorized Test & Measurement Supplier
DPO4000 Aux In has 1 MΩ input impedance; a TekVPI active probe triggers the warning. Clear it and use the probe, but double the trigger level.
Why does an 'Incompatible Probe' message appear when inserting a probe into AUX In?
The DPO4000 Aux In has a TekVPI connector that uses a TPA-BNC (terminating to 1 MΩ) to power TekVPI or TekProbe BNC accessories. Because the Aux Input impedance is 1 MΩ, connecting a TekVPI active probe (e.g. TAP1500) produces an 'incompatible probe' message. You can clear the message and still use the probe. However, TekVPI probes (TAP1500) and TekProbe probes (P6245 with TPA-BNC) are expected to terminate into 50 Ω, so the signal reaching the scope will be twice the expected amplitude — set the trigger level to twice the expected value.