Wanbo Group Instruments · Authorized Test & Measurement Supplier
Determine first whether the relevant safety standard requires AC or DC test voltage. When both are allowed, manufacturers should understand the pros and cons of each test voltage.
First confirm whether the relevant safety authority allows AC or DC test voltage for the product under test, and test accordingly. When both are permitted, manufacturers may choose based on product characteristics, but users must understand the advantages and disadvantages of each.
Most DUTs have some stray capacitance; during AC testing a continuous capacitive current flows through it.
Advantages: AC testing is generally more readily accepted by safety authorities because most products run on AC; it tests both polarities simultaneously and matches real use. Unless the product is especially sensitive to surge or the standard requires slow ramp-up, full voltage can be applied immediately. No discharge wait is usually needed after the test.
Disadvantages: if the DUT has large stray capacitance or is a capacitive load, the capacitive current far exceeds the real leakage current, so true leakage cannot be measured. The instrument must supply the capacitive current, requiring higher power than DC testing.
During DC testing the stray capacitor charges fully and the capacitive current eventually drops to zero.
Advantages: once charged, only the true leakage current remains, so DC testing clearly reveals actual leakage. Since charging current is only needed briefly, the required instrument power is far lower than for AC testing.
Disadvantages: unless there is no capacitance at all, the voltage must ramp slowly from zero to avoid excessive charging current; larger capacitance needs longer ramp-up and smaller step voltage, otherwise false rejection occurs. The DUT must be discharged before further work. DC testing is single-polarity only, which must be considered for AC-powered products. Finally, since AC peak is 1.4x the indicated RMS, safety authorities require the DC test voltage to be raised to an equivalent value.