An open probe is noise-prone because its impedance approaches infinity; place a grounded metal plate under the probe (or use the meter top) to improve success.
Q: I tried to run open compensation on my LCR meter by selecting ALL, but an error appeared midway and I could not complete the procedure. What is the issue?
A: Conduct open compensation by placing a grounded metal plate under the probe. You can also use the top of the LCR meter as a substitute for the metal plate. An open probe is more susceptible to noise because its impedance approaches infinity. Reducing noise with a grounded metal plate helps open compensation succeed.