Ensure the DUT is safely grounded and de-energized, the E terminal is grounded, and the battery is charged. Touch the start/stop key to apply high voltage and to turn it off.
Ensure the DUT is safely grounded and de-energized.
Ensure the meter's E (ground) terminal is grounded.
Ensure the battery is charged; if the low-battery lamp is on, testing is not possible.
Touch the START/STOP key; high voltage appears at the E and L terminals. Be careful!
After the test, touch START/STOP again to turn off the high voltage.
1. Connect the DUT firmly with the dedicated test leads.
2. Power on: turn the power switch on; the indicator lights and the LCD shows open-circuit "1".
3. Enable high voltage: select the required test voltage with the voltage key; the corresponding lamp lights. Press START/STOP and the instrument applies high voltage.
4. Insulation resistance test: connect E to the DUT's ground terminal and L to its line terminal. Start high voltage; the screen shows the insulation resistance value.
5. Turn off high voltage: when finished, press START/STOP; the lamp goes out and high voltage is off.
6. Charging: connect AC220V to the charging port; the charge lamp lights and turns off when full. If stored long-term, recharge every 3-5 months.