Choose the surge method by cable type: unbalanced lines use an asymmetrical CDN for line-line/line-ground tests, balanced lines use a symmetrical CDN for line-ground tests, with notes on coupling-device waveform effects and calibration.
Choose the method by the signal-port cable type. Unshielded unbalanced lines require line-line and line-ground tests with an asymmetrical CDN; unshielded balanced lines require all-lines-ground tests with a symmetrical CDN. For shielded lines, when both ends are grounded, an 18µF capacitor is inserted at the generator output to inject the pulse directly onto the shield.
Each line has an independent decoupling inductor, enabling both line-line differential and line-ground common-mode tests. It supports 1.2/50µs combination waves and 10/700µs waves (special). Pay attention to how coupling devices (capacitors, TVS, GDT) affect the waveform.
A symmetrical CDN groups two lines per common-mode inductor and supports only all-lines-ground common-mode tests; line-line differential tests pair group 1/2 with 3/4. For surge waves the decoupling network must present high LR impedance. High-speed networks suit only simultaneous all-lines-ground common-mode tests, which reduce winding length and line impedance and attenuate the EUT's high-speed signal less.
The test waveform is mainly the 1.2/50µs (8/20µs) combination wave. Open-circuit voltage is measured with a differential probe and short-circuit current with a current probe. During CDN waveform measurement, all AE-port lines are shorted to ground to assess inductor magnetic saturation, decoupling ability and waveform impact.