Wanbo Group Instruments · Authorized Test & Measurement Supplier
In SIGLENT specs, div equals the vertical scale value (e.g. V/div). Example using the SDS1052DL+ DC gain accuracy formula.
Q: What does "div" mean in the specifications?
A: For most SIGLENT specifications, "div" equals the Vertical Scale value (commonly shown in Volts/div). Take the DC Gain Accuracy of a SIGLENT SDS1052DL+. For ranges ≤ 100 mV/div, the DC Gain Accuracy is:
± [3% × (|reading| + |offset|) + 1% × |offset| + 0.2 × div + 2 mV]
Example: Reading = 0.5 V, Offset = 0.1 V, Vertical scale (div) = 100 mV/div = 0.1 V/div:
Total = 0.018 + 0.0001 + 0.02 + 0.002 ≈ 0.0401 V