Wanbo Group Instruments · Authorized Test & Measurement Supplier
Using the quadrant source-measure unit (SMU) to fully characterize a Si PIN photodiode: dark current, responsivity and breakdown voltage.
The Source Measure Unit (SMU) is the core instrument for optoelectronic testing; its quadrant sourcing/measurement fits the bidirectional energy characteristics of photo-devices across R&D and production.
Example: a Si PIN photodiode for small photovoltaic detectors; the goal is to verify dark current, responsivity and breakdown voltage for low leakage and high efficiency.
Set reverse-bias sweep 0-10V, step 0.5V, fA-level current resolution; enable over-voltage (12V) and over-current (10µA) protection.
With the light source off, ramp reverse bias 0 to 10V, hold 1s per step, and sample reverse leakage (dark current) at fA resolution.
Under stable illumination, continue the bias sweep and measure photocurrent to compute responsivity and plot the I-V curve. For breakdown, step up bias slowly and record the voltage where current rises abruptly.