How LCR meters derive L/C/R from Z and θ, choosing series/parallel equivalent circuits, open/short correction, and signal-level modes.
The impedance Z consists of a real part (Rs) and an imaginary part (X): Z = Rs + jX, |Z| = √(Rs²+X²), θ = tan⁻¹(X/Rs). LCR meters measure current through the target and voltage across its terminals to derive Z and θ, then calculate L, C, and R.
The instrument cannot determine the right mode itself. Series mode (Cs/Ls in series with Rs) suits low-impedance elements (~100 Ω or less), such as large capacitors and low inductances. Parallel mode (Cp/Lp in parallel with Rp) suits high-impedance elements (~10 kΩ or more), such as small capacitors and high inductances.
The fixture has residual components, so the measured value Zm contains residual impedance and admittance. Open and short correction remove these residuals to yield the true value Zx; LCR meters include both functions.
Three modes exist: open-voltage (V) mode sets the open-terminal voltage; constant-voltage (CV) mode sets Vx across the target (for voltage-dependent devices like high-k MLCCs); constant-current (CC) mode holds a fixed current.