Using 4/5/6 Series MSOs and PWR software: probe deskew, power quality, harmonics, inrush, switching loss, RDS(on), SOA and magnetic analysis.
From portable devices to industrial and energy systems, designers must evaluate steady-state performance as well as switching loss, control-loop stability and transient response. This note covers using Tektronix 4/5/6 Series B MSOs with 4-PWR/5-PWR/6-PWR software.
Compensate voltage/current probe skew, remove probe offset and degauss current probes. The MSO offers one-button static deskew: it reads nominal probe propagation delays and aligns the relative timing at the push of a button.
Power-quality measurements (frequency, RMS V/I, crest factor, real/reactive/apparent power, power factor and phase angle) use a differential voltage probe and a current probe on the AC line.
Non-linear loads draw current at multiples of the fundamental. Standards like IEC 61000-3-2, MIL-STD-1399 and DO-160G limit harmonics; the application displays tabular/bar results and compares against pre-defined limits for pre-compliance checks.
At power-on, input capacitors charge and draw high inrush current. The software auto-detects the inrush region and calculates the inrush current.
The software auto-measures inductance, magnetic loss and B-H, and separates line-frequency ripple from switching ripple with selectable coupling, bandwidth limit and acquisition mode.