Steps to establish communication between a VT311 and PC
FAQVTECHSeptember 18, 2026
How to connect the VT311 chassis over LAN in automatic (DHCP) or manual (static IP) mode.
Steps to establish VT311-to-PC communication
Answer:
Install the NI-VISA environment.
Auto (DHCP) mode: set the PC Ethernet adapter to automatic (DHCP), then manually add the LAN resource in NI-MAX.
Manual mode: set the PC IP to 192.168.1.200; in NI-MAX (Devices and Interfaces → Network Devices), add a VXI-11 device with static IP 192.168.1.100 (check the current IP on the front-panel LCD). When the resource name appears, the driver works.
See the VT311 manual for details on adding LAN devices.